ChpConfiguration
Provides methods to configure the Chp measurement.
- class nirfmxcdma2k.chp_configuration.ChpConfiguration(signal_obj)[source]
Bases:
objectProvides methods to configure the Chp measurement.
- configure_averaging(selector_string, averaging_enabled, averaging_count, averaging_type)[source]
Configures averaging for the CHP measurement.
- Parameters:
selector_string (string) – Pass an empty string. The signal name that is passed when creating the signal configuration is used.
averaging_enabled (enums.ChpAveragingEnabled, int) –
This parameter specifies whether to enable averaging for the measurement. The default value is False.
Name (value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The CHP measurement uses the Averaging Count parameter to calculate the number of acquisitions over which the CHP measurement is averaged.
averaging_count (int) – This parameter specifies the number of acquisitions used for averaging when you set the Averaging Enabled parameter to True. The default value is 10.
averaging_type (enums.ChpAveragingType, int) –
This parameter specifies the averaging type for averaging multiple spectrum acquisitions. RFmx uses the averaged spectrum for the measurement.
The default value is RMS.
Name (value)
Description
RMS (0)
The power spectrum is linearly averaged. RMS averaging reduces signal fluctuations but not the noise floor.
Log (1)
The power spectrum is averaged in a logarithmic scale.
Scalar (2)
The square root of the power spectrum is averaged.
Max (3)
The peak power in the spectrum at each frequency bin is retained from one acquisition to the next.
Min (4)
The least power in the spectrum at each frequency bin is retained from one acquisition to the next.
Refer to the Averaging section of the Spectrum topic for more information about averaging types.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- configure_rbw_filter(selector_string, rbw_auto, rbw, rbw_filter_type)[source]
Configures the RBW filter.
- Parameters:
selector_string (string) – Pass an empty string. The signal name that is passed when creating the signal configuration is used.
rbw_auto (enums.ChpRbwAutoBandwidth, int) –
This parameter specifies whether the measurement computes the RBW.
The default value is True.
Refer to the RBW and Sweep Time section in the CHP topic for more information about RBW and sweep time.
Name (value)
Description
False (0)
The measurement uses the RBW that you specify in the RBW parameter.
True (1)
The measurement computes the RBW.
rbw (float) – This parameter specifies the bandwidth of the RBW filter used to sweep the acquired signal if you set the RBW Auto parameter to False. This value is expressed in Hz.
rbw_filter_type (enums.ChpRbwFilterType, int) –
This parameter specifies the shape of the digital RBW filter. The default value is Gaussian.
Name (value)
Description
FFT Based (0)
No RBW filtering is performed.
Gaussian (1)
An RBW filter with a Gaussian response is applied.
Flat (2)
An RBW filter with a flat response is applied.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- configure_sweep_time(selector_string, sweep_time_auto, sweep_time_interval)[source]
Configures the sweep time.
- Parameters:
selector_string (string) – Pass an empty string. The signal name that is passed when creating the signal configuration is used.
sweep_time_auto (enums.ChpSweepTimeAuto, int) –
This parameter specifies whether the measurement computes the sweep time.
The default value is True .
Name (value)
Description
False (0)
The measurement uses the sweep time that you specify using the Sweep Time Interval parameter.
True (1)
The measurement uses the default sweep time of 0.001667 seconds.
Refer to the RBW and Sweep Time section in the CHP topic for more information about RBW and sweep time.
sweep_time_interval (float) –
This parameter specifies the sweep time if you set the Sweep Time Auto parameter to False. This value is expressed in seconds.
The default value is 0.00167 seconds.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- get_all_traces_enabled(selector_string)[source]
Gets whether to enable the traces to be stored and retrieved after performing the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (bool):
Specifies whether to enable the traces to be stored and retrieved after performing the CHP measurement.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_averaging_count(selector_string)[source]
Gets the number of acquisitions used for averaging when you set the
CHP_AVERAGING_ENABLEDattribute to True.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (int):
Specifies the number of acquisitions used for averaging when you set the
CHP_AVERAGING_ENABLEDattribute to True.- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_averaging_enabled(selector_string)[source]
Gets whether to enable averaging for the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The CHP measurement uses the CHP Averaging Count attribute as the number of acquisitions over which the CHP measurement is averaged.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (enums.ChpAveragingEnabled):
Specifies whether to enable averaging for the CHP measurement.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_averaging_type(selector_string)[source]
Gets the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is RMS.
Name (Value)
Description
RMS (0)
The power spectrum is linearly averaged. RMS averaging reduces signal fluctuations but not the noise floor.
Log (1)
The power spectrum is averaged on a logarithmic scale.
Scalar (2)
The square root of the power spectrum is averaged.
Max (3)
The peak power in the spectrum at each frequency bin is retained from one acquisition to the next.
Min (4)
The lowest power in the spectrum at each frequency bin is retained from one acquisition to the next.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (enums.ChpAveragingType):
Specifies the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the CHP measurement.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_integration_bandwidth(selector_string)[source]
Gets the CHP carrier integration bandwidth. This value is expressed in Hz.
You do not need to use a selector string to read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (float):
Returns the CHP carrier integration bandwidth. This value is expressed in Hz.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_measurement_enabled(selector_string)[source]
Gets whether to enable the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (bool):
Specifies whether to enable the CHP measurement.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_number_of_analysis_threads(selector_string)[source]
Gets the maximum number of threads used for parallelism for the CHP measurement.
The number of threads can range from 1 to the number of physical cores. The number of threads you set may not be used in calculations. The actual number of threads used depends on the problem size, system resources, data availability, and other considerations.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (int):
Specifies the maximum number of threads used for parallelism for the CHP measurement.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_rbw_filter_auto_bandwidth(selector_string)[source]
Gets whether the measurement computes the RBW.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is True.
Name (Value)
Description
False (0)
The measurement uses the RBW that you specify in the CHP RBW attribute.
True (1)
The measurement computes the RBW.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (enums.ChpRbwAutoBandwidth):
Specifies whether the measurement computes the RBW.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_rbw_filter_bandwidth(selector_string)[source]
Gets the bandwidth of the RBW filter used to sweep the acquired signal when you set the
CHP_RBW_FILTER_AUTO_BANDWIDTHattribute to False. This value is expressed in Hz.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10 kHz.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (float):
Specifies the bandwidth of the RBW filter used to sweep the acquired signal when you set the
CHP_RBW_FILTER_AUTO_BANDWIDTHattribute to False. This value is expressed in Hz.- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_rbw_filter_type(selector_string)[source]
Gets the shape of the digital RBW filter.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Gaussian.
Name (Value)
Description
FFT Based (0)
No RBW filtering is performed.
Gaussian (1)
An RBW filter with a Gaussian response is applied.
Flat (2)
An RBW filter with a flat response is applied.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (enums.ChpRbwFilterType):
Specifies the shape of the digital RBW filter.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_sweep_time_auto(selector_string)[source]
Gets whether the measurement computes the sweep time.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is True.
Name (Value)
Description
False (0)
The measurement uses the sweep time that you specify in the CHP Sweep Time attribute.
True (1)
The measurement uses the default sweep time of 0.001667 seconds.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (enums.ChpSweepTimeAuto):
Specifies whether the measurement computes the sweep time.
- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- get_sweep_time_interval(selector_string)[source]
Gets the sweep time when you set the
CHP_SWEEP_TIME_AUTOattribute to False. This value is expressed in seconds.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.001667 seconds.
- Parameters:
selector_string (string) – Pass an empty string.
- Returns:
- attr_val (float):
Specifies the sweep time when you set the
CHP_SWEEP_TIME_AUTOattribute to False. This value is expressed in seconds.- error_code (int):
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
Tuple (attr_val, error_code)
- set_all_traces_enabled(selector_string, value)[source]
Sets whether to enable the traces to be stored and retrieved after performing the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- Parameters:
selector_string (string) – Pass an empty string.
value (bool) – Specifies whether to enable the traces to be stored and retrieved after performing the CHP measurement.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_averaging_count(selector_string, value)[source]
Sets the number of acquisitions used for averaging when you set the
CHP_AVERAGING_ENABLEDattribute to True.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- Parameters:
selector_string (string) – Pass an empty string.
value (int) – Specifies the number of acquisitions used for averaging when you set the
CHP_AVERAGING_ENABLEDattribute to True.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_averaging_enabled(selector_string, value)[source]
Sets whether to enable averaging for the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The CHP measurement uses the CHP Averaging Count attribute as the number of acquisitions over which the CHP measurement is averaged.
- Parameters:
selector_string (string) – Pass an empty string.
value (enums.ChpAveragingEnabled, int) – Specifies whether to enable averaging for the CHP measurement.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_averaging_type(selector_string, value)[source]
Sets the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is RMS.
Name (Value)
Description
RMS (0)
The power spectrum is linearly averaged. RMS averaging reduces signal fluctuations but not the noise floor.
Log (1)
The power spectrum is averaged on a logarithmic scale.
Scalar (2)
The square root of the power spectrum is averaged.
Max (3)
The peak power in the spectrum at each frequency bin is retained from one acquisition to the next.
Min (4)
The lowest power in the spectrum at each frequency bin is retained from one acquisition to the next.
- Parameters:
selector_string (string) – Pass an empty string.
value (enums.ChpAveragingType, int) – Specifies the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the CHP measurement.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_measurement_enabled(selector_string, value)[source]
Sets whether to enable the CHP measurement.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- set_number_of_analysis_threads(selector_string, value)[source]
Sets the maximum number of threads used for parallelism for the CHP measurement.
The number of threads can range from 1 to the number of physical cores. The number of threads you set may not be used in calculations. The actual number of threads used depends on the problem size, system resources, data availability, and other considerations.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- Parameters:
selector_string (string) – Pass an empty string.
value (int) – Specifies the maximum number of threads used for parallelism for the CHP measurement.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_rbw_filter_auto_bandwidth(selector_string, value)[source]
Sets whether the measurement computes the RBW.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is True.
Name (Value)
Description
False (0)
The measurement uses the RBW that you specify in the CHP RBW attribute.
True (1)
The measurement computes the RBW.
- Parameters:
selector_string (string) – Pass an empty string.
value (enums.ChpRbwAutoBandwidth, int) – Specifies whether the measurement computes the RBW.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_rbw_filter_bandwidth(selector_string, value)[source]
Sets the bandwidth of the RBW filter used to sweep the acquired signal when you set the
CHP_RBW_FILTER_AUTO_BANDWIDTHattribute to False. This value is expressed in Hz.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10 kHz.
- Parameters:
selector_string (string) – Pass an empty string.
value (float) – Specifies the bandwidth of the RBW filter used to sweep the acquired signal when you set the
CHP_RBW_FILTER_AUTO_BANDWIDTHattribute to False. This value is expressed in Hz.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_rbw_filter_type(selector_string, value)[source]
Sets the shape of the digital RBW filter.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Gaussian.
Name (Value)
Description
FFT Based (0)
No RBW filtering is performed.
Gaussian (1)
An RBW filter with a Gaussian response is applied.
Flat (2)
An RBW filter with a flat response is applied.
- Parameters:
selector_string (string) – Pass an empty string.
value (enums.ChpRbwFilterType, int) – Specifies the shape of the digital RBW filter.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_sweep_time_auto(selector_string, value)[source]
Sets whether the measurement computes the sweep time.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is True.
Name (Value)
Description
False (0)
The measurement uses the sweep time that you specify in the CHP Sweep Time attribute.
True (1)
The measurement uses the default sweep time of 0.001667 seconds.
- Parameters:
selector_string (string) – Pass an empty string.
value (enums.ChpSweepTimeAuto, int) – Specifies whether the measurement computes the sweep time.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type:
- set_sweep_time_interval(selector_string, value)[source]
Sets the sweep time when you set the
CHP_SWEEP_TIME_AUTOattribute to False. This value is expressed in seconds.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.001667 seconds.
- Parameters:
selector_string (string) – Pass an empty string.
value (float) – Specifies the sweep time when you set the
CHP_SWEEP_TIME_AUTOattribute to False. This value is expressed in seconds.
- Returns:
Returns the status code of this method. The status code either indicates success or describes a warning condition.
- Return type: